The device consists of internal and external surfaces made from stainless steel
The door consists of heat-resistant double glass, and can be made from stainless steel optionally
Walls and doors of the device have double layers, and the wall thickness of walls vary between 20 and 25mm
depending on the thickness of the material
Air circulation is distributed homogenously with a fan
It is equipped with 4 types of user-friendly control systems selected within the ordering process
The upper and single divisions contain the key for switching between I.V. fluids and IRA/Blankets and control to
keep the temperature in the range of 32 -44°C or 32 – 71.1°C. The temperature selection for lower divisions is
between 32 and 71.1°C
Any error occurring in the device is shown on the digital screen
The device will produce audiovisual warning when the set temperature exceeds 5.5°C (± 0.5 ° C) and turns the
heater off automatically
The double cabin control system controls the cabins and the temperatures of both cabins. Upper and lower
cabins can operate at different temperatures